FDA recall record

Covidien Capnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01 recall — FDA Z-0515-2021

The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature…

Class IIclassification
Z-0515-2021official record ID
Dec 9, 2020report date

Recall record at a glance

Official recordZ-0515-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCovidien LLC
Report dateDec 9, 2020
Recall initiation dateNov 16, 2020
Quantity3217
LocationMansfield, MA, United States

Official product description

Capnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01, PM35MN05, DLPM35MN02 - Product Usage: use during both no motion and motion conditions and for patients who are well or poorly perfused.

Why the product was recalled

The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm.

Distribution information

Worldwide distribution - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.