FDA recall record

Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 recall — FDA Z-0198-2023

After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an…

Class IIclassification
Z-0198-2023official record ID
Nov 16, 2022report date

Recall record at a glance

Official recordZ-0198-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCanon Medical System, USA, Inc.
Report dateNov 16, 2022
Recall initiation dateSep 29, 2022
Quantity65
LocationTustin, CA, United States

Official product description

Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450

Why the product was recalled

After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.

Distribution information

US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.