FDA recall record

Trumpf Medical Systems Camera System, Television, Surgical without audio Intended for recall — FDA Z-0824-2018

The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other…

Class IIclassification
Z-0824-2018official record ID
Mar 7, 2018report date

Recall record at a glance

Official recordZ-0824-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTrumpf Medical Systems, Inc.
Report dateMar 7, 2018
Recall initiation dateDec 15, 2017
Quantity212 devices
LocationCharleston, SC, United States

Official product description

Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.

Why the product was recalled

The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.

Distribution information

Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Poland, Portugal, Romania, Sweden, Slovenia, Thailand, Turkey.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.