Recall record at a glance
| Official record | Z-1284-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | CAIRE DIAGNOSTICS Inc. |
| Report date | Mar 31, 2021 |
| Recall initiation date | Feb 26, 2021 |
| Quantity | 381 units (171 units U.S. and 210 units O.U.S.) |
| Location | Pleasanton, CA, United States |
Official product description
CAIRE DIAGNOSTICS FenomPro REF 900-0001 SN ***** - Product Usage: a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.
Why the product was recalled
Erroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease. Two issues affecting the FeNO readings: 1) Drift in the calibration gases and 2) Software error. Each of these issues can cause FeNO scores/results to be erroneously high.
Distribution information
Worldwide distribution - U.S. Nationwide distribution in the states of CA, CO, FL, TX, and WI. In the countries of Belgium, Canada, Croatia/Hrvatska, France, Germany, Greece, Hong Kong, India, Philippines, Spain, Spain and Switzerland.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.