FDA recall record

Smiths Medical ASD CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder recall — FDA Z-1439-2017

20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key…

Class IIclassification
Z-1439-2017official record ID
Mar 22, 2017report date

Recall record at a glance

Official recordZ-1439-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSmiths Medical ASD, Inc.
Report dateMar 22, 2017
Recall initiation dateOct 31, 2016
Quantity20
LocationSaint Paul, MN, United States

Official product description

CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15,

Why the product was recalled

20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect message. The message indicates that a Patient Controlled Analgesia (PCA) dose is unavailable because the pump is running. It should indicate that the PCA dose is not available because the pump is stopped. The function of the pump is unchanged and no patient injury can occur since no drug is being delivered.

Distribution information

Internationally to Finland

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.