FDA recall record

Smiths Medical ASD CADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF recall — FDA Z-0166-2015

Smiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs.…

Class IIclassification
Z-0166-2015official record ID
Nov 12, 2014report date

Recall record at a glance

Official recordZ-0166-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSmiths Medical ASD, Inc.
Report dateNov 12, 2014
Recall initiation dateSep 29, 2014
Quantity24
LocationSaint Paul, MN, United States

Official product description

CADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF 21-2194-0301-01. Product Usage: The CADD-Solis Medication Safety Software- Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD-Solis VIP Ambulatory Infusion Pump, CADD-Solis Ambulatory Infusion Pump, or CADD-Prizm PCSII Ambulatory Infusion Pump (software revision H or higher). The CADD-Solis Medication Safety Software- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD-Solis Medication Safety Software.- Administrator to the CADD -Solis Ambulatory Infusion Pump and CADD-Prizm PCS I I Ambulatory Infusion Pump (software revision H or higher).

Why the product was recalled

Smiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs. CADD"-Solis Medication Safety Software Administrator 3.1 CDs were sent with the CADD"-Solis Medication Safety Software Point of Care 3.1 software loaded on them.

Distribution information

US Nationwide Distribution in the state of AK, CA, CO, CT, GA, MI, MN, NH, NC, SC, TX, VA, WI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.