FDA recall record

Smiths Medical ASD CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are recall — FDA Z-0222-2025

Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant,…

Class Iclassification
Z-0222-2025official record ID
Nov 6, 2024report date

Recall record at a glance

Official recordZ-0222-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmSmiths Medical ASD Inc.
Report dateNov 6, 2024
Recall initiation dateJul 16, 2024
Quantity138039 units
LocationMinneapolis, MN, United States

Official product description

CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.

Why the product was recalled

Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

Distribution information

Worldwide Distribution

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.