FDA recall record

Smiths Medical ASD CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0 recall — FDA Z-1813-2015

Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the…

Class IIclassification
Z-1813-2015official record ID
Jun 24, 2015report date

Recall record at a glance

Official recordZ-1813-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSmiths Medical ASD, Inc.
Report dateJun 24, 2015
Recall initiation dateMay 26, 2015
Quantity4426 (2427 US, 1999 OUS)
LocationSaint Paul, MN, United States

Official product description

CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 & Model 2120, Version 1.0. Provide measured drug therapy to patients in hospital or outpatient settings.

Why the product was recalled

Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD¿-Solis Medication Cassette Reservoir or Administration Set disposable to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (e

Distribution information

Worldwide Distribution-US (nationwide) including the states of AL, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, LA, MA, MI, MN, MS, MO, NE, NM, NY, NC, ND, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, and WI, and the countries of AT (Austria), AU (Australia), BE (Belgium), CA (Canada), CH (Switzerland), DE (Germany), DK (Denmark), FR (France), GB (Great Britain), HK (Hong Kong), LB (Lebanon), NL (Netherlands), NZ (New Zealand), QA (Qatar), SA (Saudi Arabia), and SE (Sweden).

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.