FDA recall record

Smiths Medical ASD CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01 recall — FDA Z-1831-2016

Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.

Class IIclassification
Z-1831-2016official record ID
Jun 8, 2016report date

Recall record at a glance

Official recordZ-1831-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSmiths Medical ASD, Inc.
Report dateJun 8, 2016
Recall initiation dateApr 22, 2016
Quantity13,696,830 units
LocationSaint Paul, MN, United States

Official product description

CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01, 21-7321-24, 21-7322-01, 21-7322-24, 21-7323-24, 21-7324-01, 21-7324-24, 21-7333-24, 21-7336-01, 21-7336-24, 21-7339-01, 21-7339-24, 21-7359-01, 21-7359-24, 21-7383-01, 21-7383-24, 21-7390-01, 21-7390-24, 21-7391-01, 21-7391-24, 21-7394-01, 21-7394-24, 21-7395-24. The CADD Medication Cassette Reservoirs with Flow Stop are a modification to the current CADD Medication Cassette Reservoirs. The reservoirs will incorporate a set-based free flow protection component (i.e. Flow Stop) that is designed to occlude the tube if the reservoirs are accidentally placed onto the pump incorrectly or become detached from the pump. The Flow Stop will be located on the reservoir housing, which is attached to the pump. The reservoir will be provided to the user in an open state. Before the reservoir can be attached to the pump, the blue "CLIP" must be removed to activate the Flow Stop. However, after attaching the reservoir to the pump, the user can still remove the reservoir from the pump and prime it by holding the Flow Stop in the open position.

Why the product was recalled

Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.

Distribution information

US: ALL STATES IN CONTINENTAL USA INCLUDING DC AND PR EXCEPT DE. OUS: UNITED ARAB EMIRATES, AUSTRIA BELGIUM, CANADA, CHINA, GERMANY, DENMARK, SPAIN, FINLAND, FRANCE, GREAT BRITAIN, HONG KONG, IRELAND, INDIA, ITALY, JAPAN, LUXEMBOURG, MAURITIUS, NEW CALEDONIA, NEW ZEALAND, PANAMA, PORTUGAL, SAUDI ARABIA, SWEDEN, SINGAPORE, TAIWAN, VIET NAAM, SOUTH AFRICA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records