FDA recall record

RTI Surgical C-Plus" 10mm x 12mm Webless implants Pioneer Surgical's C-Plus " 10mm x 12mm recall — FDA Z-0627-2014

As the result of a recent internal review of regulatory documents, we have determined that the 10 mm x 12 mm Webless C-Plus implants were incorrectly…

Class IIclassification
Z-0627-2014official record ID
Jan 15, 2014report date

Recall record at a glance

Official recordZ-0627-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRTI Surgical, Inc.
Report dateJan 15, 2014
Recall initiation dateOct 7, 2013
Quantity1906
LocationMarquette, MI, United States

Official product description

C-Plus" 10mm x 12mm Webless implants Pioneer Surgical's C-Plus " 10mm x 12mm Webless implants are cervical intervertebral fusion devices that are intended to stabilize a cervical spinal segment to promote fusion in order to restrict motion and decrease pain. They are intended for use with supplemental fixation and autograft. These implants are manufactured from PEEK (Polyaryletheretherketone) polymer with tantalum marker pins for radiographic visualization.

Why the product was recalled

As the result of a recent internal review of regulatory documents, we have determined that the 10 mm x 12 mm Webless C-Plus implants were incorrectly documented as meeting the requirement for labeling as a partial Vertebral Body Replacement Device (VBR). While the use of this device as a partial VBR is unlikely due to the size of the device, use of this device as a partial VBR could result in a s

Distribution information

Nationwide Distribution - USA including FL, MO, NY, CA, MI,and IL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.