FDA recall record

Pacific Medical Group Button-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700HAX recall — FDA Z-1161-2020

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or…

Class IIclassification
Z-1161-2020official record ID
Feb 19, 2020report date

Recall record at a glance

Official recordZ-1161-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPacific Medical Group Inc.
Report dateFeb 19, 2020
Recall initiation dateSep 11, 2019
Quantity6481
LocationSan Clemente, CA, United States

Official product description

Button-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE Corometrics Nautilus Ultrasound Cable Assembly, NFCM9520; Pacific Medical GE Corometrics Nautilus Ultrasound Crystal, NFCM9535

Why the product was recalled

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.

Distribution information

Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records