FDA recall record

Philips BrightView, gamma camera system recall — FDA Z-0929-2024

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion,…

Class Iclassification
Z-0929-2024official record ID
Feb 14, 2024report date

Recall record at a glance

Official recordZ-0929-2024
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmPhilips North America
Report dateFeb 14, 2024
Recall initiation dateDec 15, 2023
Quantity302 US; 264 OUS
LocationCambridge, MA, United States

Official product description

BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A

Why the product was recalled

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Distribution information

Domestic distribution nationwide. International distribution worldwide.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.