Synaptive Medical BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and recall — FDA Z-0470-2019

This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain…

Class Iclassification
Z-0470-2019official record ID
Dec 5, 2018report date

Recall record at a glance

Official recordZ-0470-2019
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmSynaptive Medical Inc.
Report dateDec 5, 2018
Recall initiation dateOct 9, 2018
Quantity37 units
LocationToronto, N/A, Canada

Official product description

BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.

Why the product was recalled

This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.

Distribution information

Worldwide - US Nationwide Distribution and the country of Canada

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.