Recall record at a glance
| Official record | Z-2861-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Nuclein LLC |
| Report date | Aug 12, 2026 |
| Recall initiation date | Jun 25, 2025 |
| Quantity | 2 |
| Location | Austin, TX, United States |
Official product description
Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.
Why the product was recalled
Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).
Distribution information
US Nationwide distribution in the state of Texas.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.