BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO recall — FDA Z-0172-2018

Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the…

Class IIclassification
Z-0172-2018official record ID
Dec 6, 2017report date

Recall record at a glance

Official recordZ-0172-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBrainlab AG
Report dateDec 6, 2017
Recall initiation dateNov 3, 2017
Quantity28 units
LocationMunich, N/A, Germany

Official product description

BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW UPDATE 3.5.2 TO 3.5.3 46218 EXACTRAC VERO SW UPDATE 3.5.3 TO 3.5.4 The ExacTrac Vero system is intended to be used in conjunction with the MHI-TM2000 Radiation Therapy Linear Accelerator System manufactured by Mitsubishi Heavy Industries, Ltd. ExacTrac Vero uses the images received from the MHI-TM2000 linear accelerator for analyzing the current patient position and calculating - when applicable - a necessary correction shift. The correction shift is then exported to the MHI-TM2000 linear accelerator. The ExacTrac Vero system uses stereoscopic X-ray or Cone Beam CT registration and optical tracking of infrared reflective markers in order to localize and correct the patient position before and during treatment. Optionally ExacTrac Vero provides position data for the pan/tilt motion of the MHI-TM2000 gantry head to the MHITM2000 controller for continuous alignment of the beam orientation with a moving target. The position data is based on target detection via X-ray imaging and IR tracking of external surrogate markers.

Why the product was recalled

Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.

Distribution information

Worldwide Distribution - US Distribution and to the countries of : Belgium, Canada, France, Germany, Italy, Japan and S. Korea

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 17, 2021

Brainlab AG ExacTrac Dynamic recall — FDA Z-1185-2021

In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may…

Class IIFDA: TerminatedBrainlab AG