FDA recall record

Braemar Manufacturing LLC DL950 Holter Monitor 1285 Corporate Center Dr recall — FDA Z-1460-2020

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally.…

Class IIclassification
Z-1460-2020official record ID
Mar 18, 2020report date

Recall record at a glance

Official recordZ-1460-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBraemar Manufacturing, LLC
Report dateMar 18, 2020
Recall initiation dateJan 22, 2020
Quantity10582
LocationEagan, MN, United States

Official product description

Braemar Manufacturing LLC DL950 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only

Why the product was recalled

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

Distribution information

Nationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.