Recall record at a glance
| Official record | Z-0824-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Zimmer Manufacturing B.V. |
| Report date | Dec 24, 2014 |
| Recall initiation date | Nov 21, 2014 |
| Quantity | 6,377 |
| Location | N/A |
Official product description
Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and cluster-holed. The multi and cluster-holed shells allow for the use of screws to provide additional fixation and security, particularly in those case where acetabular bone stock is deficient. The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Why the product was recalled
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .
Distribution information
Worldwide Distribution - USA (nationwide) and Internationally to Canada Australia Brazil Chile China Colombia Japan Malaysia Singapore South Korea Taiwan Albania Austria Belgium Bulgaria Canary Islands Czech Republic Denmark Egypt Finland France Germany Greece Hungary Ireland Italy Netherlands Norway Poland Portugal Russia Saudi Arabia Serbia Slovakia South Africa Spain Sweden Switzerland Turkey and United Kingdom.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.