FDA recall record

CME BodyGuard Microset, REF: A120-161XYBS recall — FDA Z-2563-2020

Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system…

Class Iclassification
Z-2563-2020official record ID
Jul 22, 2020report date

Recall record at a glance

Official recordZ-2563-2020
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmCME America, LLC
Report dateJul 22, 2020
Recall initiation dateJun 16, 2020
Quantity18520
LocationGolden, CO, United States

Official product description

BodyGuard Microset, REF: A120-161XYBS

Why the product was recalled

Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy

Distribution information

U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.