FDA recall record

Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 recall — FDA Z-0452-2022

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have…

Class IIclassification
Z-0452-2022official record ID
Jan 12, 2022report date

Recall record at a glance

Official recordZ-0452-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBiomeme, Inc.
Report dateJan 12, 2022
Recall initiation dateOct 20, 2021
Quantity11,607
LocationPhiladelphia, PA, United States

Official product description

Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564

Why the product was recalled

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).

Distribution information

Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records