FDA recall record

Biomedix SELEC-3 I.V recall — FDA Z-0023-2015

Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix…

Class IIclassification
Z-0023-2015official record ID
Oct 15, 2014report date

Recall record at a glance

Official recordZ-0023-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBiomedix, Inc.
Report dateOct 15, 2014
Recall initiation dateAug 14, 2014
Quantity2,928
LocationBloomington, IN, United States

Official product description

Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE

Why the product was recalled

Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.

Distribution information

US Distribution including the states of OH, NC, and TX

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.