FDA recall record

Angiodynamics BioFlo Port with ENDEXO and PASV Technology-AngioDynamics BIOFLO 8F SINGLE recall — FDA Z-0076-2019

BioFlo Ports was partially manufactured with the incorrect sterile port assembly. The actual sterile port assembly used was, 8F SL Xcela Plus Titanium…

Class IIclassification
Z-0076-2019official record ID
Oct 17, 2018report date

Recall record at a glance

Official recordZ-0076-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAngiodynamics Inc. (Navilyst Medical Inc.)
Report dateOct 17, 2018
Recall initiation dateJul 24, 2018
Quantity1 box
LocationGlens Falls, NY, United States

Official product description

BioFlo Port with ENDEXO and PASV Technology-AngioDynamics BIOFLO 8F SINGLE PLASTIC NON-FILLED; VALVED PG UPN: H965440210 Catalog Number: 44-021 Box of 1 Port - Product Usage - Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

Why the product was recalled

BioFlo Ports was partially manufactured with the incorrect sterile port assembly. The actual sterile port assembly used was, 8F SL Xcela Plus Titanium Valved Non-Filled, when it should have been, 8F SL BioFlo Plastic Valved Non-Filled

Distribution information

US Nationwide in the state of IN.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.