FDA recall record

BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " recall — FDA Z-1333-2021

Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range…

Class IIclassification
Z-1333-2021official record ID
Apr 7, 2021report date

Recall record at a glance

Official recordZ-1333-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBiocare Medical, LLC
Report dateApr 7, 2021
Recall initiation dateFeb 25, 2021
Quantity72 units
LocationPacheco, CA, United States

Official product description

BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1

Why the product was recalled

Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.

Distribution information

US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records