FDA recall record

Ventana Medical Systems Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS recall — FDA Z-2118-2013

Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element…

Class IIclassification
Z-2118-2013official record ID
Sep 4, 2013report date

Recall record at a glance

Official recordZ-2118-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVentana Medical Systems Inc.
Report dateSep 4, 2013
Recall initiation dateJun 12, 2008
Quantity1217 units total
LocationTucson, AZ, United States

Official product description

Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark product lines are intended to automatically stain histological or cytological specimens on microscope slides with specific immunohistochemistry or in situ hybridization reagents for in vitro diagnostic use. The Discovery Systems are research use only instruments with similar functionality.

Why the product was recalled

Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.

Distribution information

Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.