Recall record at a glance
| Official record | Z-1804-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Beckman Coulter Inc. |
| Report date | May 23, 2018 |
| Recall initiation date | Mar 26, 2018 |
| Quantity | 523 boxes |
| Location | Brea, CA, United States |
Official product description
BECKMAN COULTER MicroScan MICroSTREP Plus Panel, MicroScan Mueller-Hinton Broth with 3% lysed horse blood, Catalog Number 81015 MicroScan Mueller Hinton Broth with 3% Lysed Horse Blood is used for inoculation of MicroScan MICroSTREP plus and MicroFAST panels. MICroSTREP plus panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae.
Why the product was recalled
Beckman Coulter has received customer complaints of lowered minimum inhibitory concentration (MIC) for antimicrobial agents Cefaclor (Ctr) and Penicillin (P) with control organism Streptococcus pneumoniae (ATCC 49619) during quality control (QC) testing on MICroSTREP plus(R) Panel Type 1 hydrated with Mueller-Hinton Broth with 3% lysed horse blood part number B1015-25 lot number 1005986140.
Distribution information
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Canada, Kuwait, South Korea and Saudi Arabia.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.