FDA recall record

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and recall — FDA Z-0093-2025

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false…

Class IIIclassification
Z-0093-2025official record ID
Oct 23, 2024report date

Recall record at a glance

Official recordZ-0093-2025
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusOngoing
Recalling firmBeckman Coulter, Inc.
Report dateOct 23, 2024
Recall initiation dateJul 16, 2024
Quantity5589 units
LocationMiami, FL, United States

Official product description

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Why the product was recalled

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.

Distribution information

worldwide

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.