FDA recall record

BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167 recall — FDA Z-1643-2018

Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers.…

Class IIclassification
Z-1643-2018official record ID
May 16, 2018report date

Recall record at a glance

Official recordZ-1643-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateMay 16, 2018
Recall initiation dateMar 2, 2018
Quantity2225 units
LocationBrea, CA, United States

Official product description

BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167, 800-7719, 800-7720, 800- 7106, 800-3049, 800-3050, 800-3079, 800-3080, 800-7190, 800-7713, 800-7714, 800-7715, 700- 7176-001, 700-7177-001 Product Usage: The iChemVELOCITY automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterade, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChemVELOCITY strips are intended for use only with the iChemVELOCITY analyzer. In particular they are not intended for visual reading. The iChemVELOCITY is not intended to be used as a Point of Care (POC) analyzer. The measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver finction. Tests performed using the iChemVELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.

Why the product was recalled

Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.

Distribution information

Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.