FDA recall record

Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 recall — FDA Z-2240-2020

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six…

Class IIclassification
Z-2240-2020official record ID
Jun 10, 2020report date

Recall record at a glance

Official recordZ-2240-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateJun 10, 2020
Recall initiation dateApr 30, 2019
Quantity95,453 units
LocationBrea, CA, United States

Official product description

Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

Why the product was recalled

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

Distribution information

US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusett, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico New York, North Carolina, North Dakota, Ohio, Oklahoma Oregon, Pennsylvania, Rhode Island, South Carolina,South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, and Wyoming. US Territories: American Samoa, Guam, and Puerto Rico. OUS Distribution: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Djibouti, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liberia, Lithuania, Malawi, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, San Marino, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Zambia, and Zimbabwe.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.