FDA recall record

Becton Dickinson & Company BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 recall — FDA Z-0095-2023

The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography…

Class IIclassification
Z-0095-2023official record ID
Oct 26, 2022report date

Recall record at a glance

Official recordZ-0095-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBecton Dickinson & Company
Report dateOct 26, 2022
Recall initiation dateSep 15, 2022
Quantity2,196,000
LocationFranklin Lakes, NJ, United States

Official product description

BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting, separating, and processing blood for blood alcohol determination.

Why the product was recalled

The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in methanol testing.

Distribution information

Distribution US nationwide, Canada, New Zealand, and Hong Kong.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.