Recall record at a glance
| Official record | Z-1142-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Terminated |
| Recalling firm | Becton Dickinson & Company |
| Report date | Mar 28, 2018 |
| Recall initiation date | Mar 22, 2018 |
| Quantity | 117,816,603 |
| Location | Franklin Lakes, NJ, United States |
Official product description
BD Vacutainer Lithium Heparin Green Top Tube; Catalog Numbers: 1. 366664 (UDI: 50382903666640); 2. 367884 (UDI: 50382903678841); 3. 367886 (UDI: 50382903678865); 4. 368040 (UDI: 50382903680400); 5. 366667 (UDI: 50382903666671); 6. 367880 (UDI: 50382903678803)
Why the product was recalled
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
Distribution information
Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.