FDA recall record

Bard Peripheral Vascular BD Recanalization System, REF: BDRECANSYSTEM recall — FDA Z-0676-2024

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute…

Class IIclassification
Z-0676-2024official record ID
Jan 17, 2024report date

Recall record at a glance

Official recordZ-0676-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBard Peripheral Vascular Inc.
Report dateJan 17, 2024
Recall initiation dateNov 14, 2023
Quantity403
LocationTempe, AZ, United States

Official product description

BD Recanalization System, REF: BDRECANSYSTEM

Why the product was recalled

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

Distribution information

US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.