Recall record at a glance
| Official record | Z-0093-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | CareFusion 303, Inc. |
| Report date | Oct 18, 2023 |
| Recall initiation date | Sep 6, 2023 |
| Quantity | 61 |
| Location | San Diego, CA, United States |
Official product description
BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01
Why the product was recalled
When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.
Distribution information
US Nationwide distribution in the states of LA, NC, NY, MI, MA, NV, KS, IA, CA.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.