FDA recall record

Becton Dickinson & Company BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive recall — FDA Z-1967-2019

The products have been confirmed to have reduced or no additive within the tube reservoir. The tubes may develop visible clots within the tube samples or…

Class IIclassification
Z-1967-2019official record ID
Jul 24, 2019report date

Recall record at a glance

Official recordZ-1967-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBecton Dickinson & Company
Report dateJul 24, 2019
Recall initiation dateJun 3, 2019
Quantity1707804
LocationFranklin Lakes, NJ, United States

Official product description

BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975 Product Usage: The Microtainer Brand Tube with EDTA(K2) Reorder No. 365974, is used to collect, anticoagulate, transport and store skin puncture blood specimens for testing the following hematological parameters: WBC, RBC ,HGB, HCT, MCV, MCH, MCHC, Platelets, 5 part differential including : Neutrophils, Lymphocytes, Monocytes, Eosinophils, Basophils, Manual WBC Differentials, Reticulocytes, and Spun Hematocrit.

Why the product was recalled

The products have been confirmed to have reduced or no additive within the tube reservoir. The tubes may develop visible clots within the tube samples or micro clots that are not easily detected during visual inspection of the tubes. A clotted sample may lead to recollection of samples or, retesting of patients, resulting in delayed reporting of test results and patient treatment.

Distribution information

The products were distributed to the following US states: AL, CA, CT, FL, GA, IL, KY, MD, MN, MO, OH, PA, TN, TX, VA, and Puerto Rico.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.