FDA recall record

Becton Dickinson & Co BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364 recall — FDA Z-0312-2019

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the…

Class IIclassification
Z-0312-2019official record ID
Nov 7, 2018report date

Recall record at a glance

Official recordZ-0312-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBecton Dickinson & Co.
Report dateNov 7, 2018
Recall initiation dateSep 24, 2018
Quantity23
LocationSparks, MD, United States

Official product description

BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Why the product was recalled

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

Distribution information

Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. Countries of Spain, Germany, France, Austria, Denmark, Switzerland, Sweden, Italy, Kuwait, Oman, Saudi Arabia, and Turkey.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.