FDA recall record

BD SWITZERLAND SARL BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD recall — FDA Z-2514-2026

Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening…

Class IIclassification
Z-2514-2026official record ID
Jun 24, 2026report date

Recall record at a glance

Official recordZ-2514-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBD SWITZERLAND SARL
Report dateJun 24, 2026
Recall initiation dateMay 21, 2026
Quantity2,465,500 units
LocationEysins, Switzerland

Official product description

BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.

Why the product was recalled

Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.

Distribution information

U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.