FDA recall record

Becton Dickinson & Co BD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the isolation recall — FDA Z-1378-2021

Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours…

Class IIclassification
Z-1378-2021official record ID
Apr 14, 2021report date

Recall record at a glance

Official recordZ-1378-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBecton Dickinson & Co.
Report dateApr 14, 2021
Recall initiation dateMar 17, 2021
Quantity33660 EA
LocationSparks, MD, United States

Official product description

BD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221557

Why the product was recalled

Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment

Distribution information

Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.