FDA recall record

CareFusion 303 BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM recall — FDA Z-1217-2026

Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with…

Class IIclassification
Z-1217-2026official record ID
Feb 4, 2026report date

Recall record at a glance

Official recordZ-1217-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCareFusion 303, Inc.
Report dateFeb 4, 2026
Recall initiation dateNov 6, 2025
Quantity189 devices
LocationSan Diego, CA, United States

Official product description

BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. The BD AlarisTM System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, patient-controlled analgesia (PCA) Module, and associated software applications.

Why the product was recalled

Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network

Distribution information

U.S. Nationwide distribution in the states of AZ, IN, OH, and VA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records