FDA recall record

Baxter Novum IQ LVP INFUSION SYSTEM, REF 40700BAXUS recall — FDA Z-1767-2025

There is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered…

Class Iclassification
Z-1767-2025official record ID
May 28, 2025report date

Recall record at a glance

Official recordZ-1767-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmBaxter Healthcare Corporation
Report dateMay 28, 2025
Recall initiation dateApr 24, 2025
Quantity34,524 units
LocationDeerfield, IL, United States

Official product description

Baxter Novum IQ LVP INFUSION SYSTEM, REF 40700BAXUS

Why the product was recalled

There is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered off with the set loaded.

Distribution information

US and Canada

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Exacta Mix Vented Micro-Volume Inlet recall — FDA Z-2262-2026

Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower…

Class IIFDA: OngoingBaxter Healthcare Corporation