Recall record at a glance
| Official record | Z-1564-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Bard Peripheral Vascular Inc. |
| Report date | May 17, 2023 |
| Recall initiation date | Apr 11, 2023 |
| Quantity | 4,120 |
| Location | Tempe, AZ, United States |
Official product description
Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM
Why the product was recalled
Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.
Distribution information
US Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY . OUS: BE, CA, AU
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.