FDA recall record

Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm recall — FDA Z-0917-2020

The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the…

Class IIclassification
Z-0917-2020official record ID
Feb 5, 2020report date

Recall record at a glance

Official recordZ-0917-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBard Peripheral Vascular Inc.
Report dateFeb 5, 2020
Recall initiation dateAug 14, 2019
Quantity286 Units
LocationTempe, AZ, United States

Official product description

Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F; Nominal Pressure 8atm, Rated Burst Pressure 40atm; Syringe inflation 1cc; UDI: (01)00801741060533(17)220228(10)REDQ3444

Why the product was recalled

The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists that upon inflation, non-diseased vessel, kidney or sections of the AV dialysis fistulae may be dilated, potentially causing a varying degree of injury. This may require additional medical or surgical intervention.

Distribution information

US: NJ, NY, OH, PA, VA, SC, GA, AR, TX, MA, NC, TN, LA, AZ, MS, FL, AL,MD, CA, HI, MO, CO, WI, DC, KY, IL, MN, IN, KS, OK, CT, MT, WA, ND, IA, WA, WV, PR OUS: None

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.