FDA recall record

C.R. Bard Bard Blakemore Esophageal-Nasogastric Tube (Child) REF: 0092110 recall — FDA Z-1757-2025

Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the…

Class Iclassification
Z-1757-2025official record ID
May 28, 2025report date

Recall record at a glance

Official recordZ-1757-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmC.R. Bard Inc.
Report dateMay 28, 2025
Recall initiation dateApr 17, 2025
QuantityN/A
LocationCovington, GA, United States

Official product description

Bard Blakemore Esophageal-Nasogastric Tube (Child) REF: 0092110

Why the product was recalled

Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.

Distribution information

US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY, Puerto Rico OUS: Singapore, Canada, Belgium, Ireland, Netherlands, South Africa, Sweden, United Kingdom

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.