Recall record at a glance
| Official record | Z-0455-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | GE Healthcare Finland Oy |
| Report date | Nov 19, 2025 |
| Recall initiation date | Sep 26, 2025 |
| Quantity | 7408 |
| Location | Helsinki, Finland |
Official product description
B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms 3/19/26 updated to add B105P Patient Monitor 6160001-061 B125P Patient Monitor 6160001-062 B105M-C Patient Monitor 6160002-033 B105M-A Patient Monitor 6160000-113 B155M-A Patient Monitor 6160000-115 B125P-B Patient Monitor 6160002-022 B105M-B Patient Monitor 6160002-023 B125M-B Patient Monitor 6160002-024 B105M-OR Patient Monitor 6160000-143 B105M-OR Patient Monitor 6160000-144 B155M-OR Patient Monitor 6160000-145
Why the product was recalled
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Distribution information
Worldwide - US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.