FDA recall record

CRHF Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01 recall — FDA Z-2509-2020

There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension…

Class IIclassification
Z-2509-2020official record ID
Jul 15, 2020report date

Recall record at a glance

Official recordZ-2509-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Report dateJul 15, 2020
Recall initiation dateJun 4, 2020
Quantity31736 devices
LocationMounds View, MN, United States

Official product description

Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01

Why the product was recalled

There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.

Distribution information

worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.