FDA recall record

Linet Spol. S.r.o AVE 2 Birthing Bed recall — FDA Z-0991-2019

LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the…

Class IIclassification
Z-0991-2019official record ID
Mar 20, 2019report date

Recall record at a glance

Official recordZ-0991-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLinet Spol. S.r.o.
Report dateMar 20, 2019
Recall initiation dateJan 7, 2019
Quantity253 devices
LocationSlany, N/A, Czech Republic

Official product description

AVE 2 Birthing Bed

Why the product was recalled

LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.

Distribution information

AL, OK, NJ, OR, WA, CA, MN, LA, AZ, IL, MO, CO, OH, KY, FL, KS, UT, TN, LA, DE, CT, TX

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 13, 2021

Linet Spol. S.r.o Eleganza 5 recall — FDA Z-0767-2021

A component used within the side rail mechanism may malfunction, resulting in the side rail either not locking properly in the upper position or the side rail becoming stuck in the upper…

Class IIFDA: OngoingLinet Spol. S.r.o.