Recall record at a glance
| Official record | Z-2296-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Bayer Medical Care, Inc. |
| Report date | Jun 10, 2026 |
| Recall initiation date | Apr 24, 2026 |
| Quantity | 11,300 units (1650 US, 9650 OUS) |
| Location | Indianola, PA, United States |
Official product description
Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;
Why the product was recalled
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
Distribution information
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.