Shimadzu Corp. Analytical & Measuring Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL recall — FDA Z-0554-2025

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of…

Class IIclassification
Z-0554-2025official record ID
Dec 4, 2024report date

Recall record at a glance

Official recordZ-0554-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmShimadzu Corp. Analytical & Measuring Instruments Div.
Report dateDec 4, 2024
Recall initiation dateOct 9, 2024
Quantity48
LocationKyoto, Japan

Official product description

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Why the product was recalled

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

Distribution information

Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.