Beckman Coulter AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106 recall — FDA Z-0492-2020

Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of…

Class IIclassification
Z-0492-2020official record ID
Nov 27, 2019report date

Recall record at a glance

Official recordZ-0492-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter Inc.
Report dateNov 27, 2019
Recall initiation dateNov 9, 2018
Quantity83 automate systems
LocationBrea, CA, United States

Official product description

AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255

Why the product was recalled

Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .

Distribution information

US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.