FDA recall record

Beckman Coulter AutoMate 1200 Catalog Number ODL25120 recall — FDA Z-1169-2018

Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of…

Class IIclassification
Z-1169-2018official record ID
Mar 28, 2018report date

Recall record at a glance

Official recordZ-1169-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmBeckman Coulter Inc.
Report dateMar 28, 2018
Recall initiation dateJan 19, 2018
Quantity98 units in the US
LocationBrea, CA, United States

Official product description

AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255

Why the product was recalled

Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.

Distribution information

Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Other.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.