Recall record at a glance
| Official record | Z-2421-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Dentsply IH, Inc. |
| Report date | Sep 3, 2025 |
| Recall initiation date | Jul 24, 2025 |
| Quantity | 102 units |
| Location | Waltham, MA, United States |
Official product description
Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04B / 68020033 Software Version: n/a Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
Why the product was recalled
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
Distribution information
Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.