FDA recall record

St. Jude Medical, Cardian Rhythm Management ASSURITY Pulse Generator REF PM**** SN ********* recall — FDA Z-1530-2021

Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the…

Class Iclassification
Z-1530-2021official record ID
May 19, 2021report date

Recall record at a glance

Official recordZ-1530-2021
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmSt. Jude Medical, Cardian Rhythm Management Division
Report dateMay 19, 2021
Recall initiation dateMar 15, 2021
Quantity242,945 pacemakers [updated 3/20/2023]
LocationSylmar, CA, United States

Official product description

ASSURITY Pulse Generator REF PM**** SN ********* ST. JUDE MEDICAL Assurity and Endurity pacemakers are intended to provide single-chamber or dual-chamber bradycardia pacing.

Why the product was recalled

Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough

Distribution information

Worldwide distribution: U.S (nationwide) to states including.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.( countries): [806 Amendment - after March 2021 - updated 3/20/2023] Algeria, Angola, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Benin, Bolivia, Brazil, Bulgaria, Cameroon, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Ireland, Italy, Japan, Jordan, Korea, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Montenegro, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Palestine, Panama, Paraguay, Peru Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Senegal, Servia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records