Recall record at a glance
| Official record | Z-2344-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Aug 17, 2016 |
| Recall initiation date | Jul 7, 2016 |
| Quantity | 51 units |
| Location | Malvern, PA, United States |
Official product description
Artis zee/ zeego, Artis Q/ Q.zen, stand alone system, software controlled Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, 10848281, 10848282, 10848283, 10848353, 10848255 Artis zee / zeego and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Why the product was recalled
Due to an error in the 19 Live Display, image reproduction may fail in the examination room and the potential exists for the loss of images immediately after system startup.
Distribution information
Nationwide Distribution to AL,AZ,CA,CO,IA,IL,MA,MD,MI,MN,MO,NE,NJ,NY,OH,OR,PA,RI,TX,VA,WA,WI,WV
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.